The CELL-DYN Ruby System* is a multi-parameter automated hematology analyzer designed for in vitro diagnostic use in medium volume clinical laboratories. This instrument uses MAPSS technology to perform cell by cell analysis from a single dilution to enumerate and differentiate White Blood Cells and provide the performance needed for patient samples. Additionally, the instrument uses optical laser light scatter analysis, without the need to reflex to another test mode, to provide both the Red Blood Cell and Platelet counts.
CELL-DYN Ruby Specifications
Physical Specification
Physical specifications for the CELL-DYN Ruby are provided in the following table.Module | Height | Width | Depth | Weight |
---|---|---|---|---|
Analyzer | 49.9 cm (19.25 in.) |
86.4 cm (34.0 in.) |
76.8 cm (30.25 in.) |
105.2 kg (232.0 lbs.) |
Printers | Refer to the printer manufacturer’s specifications. |
Power Specifications
The power specifications for the CELL-DYN Ruby are described in the following tables. Refer to the power specifications applicable in your country.CELL-DYN Ruby Power Specifications
Module | Voltage | Frequency | Max Current | BTU/Hr |
---|---|---|---|---|
Analyzer | 100 - 240 VAC | 47/63 Hz | 5.0 - 2.2 amps | 550 watts |
Display | 100 - 240 VAC | 50/60 Hz | 1.5 amps | 50 watts |
Printers | For power specifications for printers, refer to the operator’s manual for your printer or other documentation received with your printer. |
Module | Fuse Rating |
---|---|
Analyzer | Internal fuses only. Not operator replaceable. |
Environmental Specifications
Environmental specifications include the operating environment required by the CELL-DYN Ruby, the clearance and waste disposal requirements, and the noise level and heat output that can be expected during normal operation.Operating Environment Requirements
Temperature : | 15°C–30°C (59°F–86°F) | |||
Relative Humidity : (non-condensing) Indoor Use |
< 80% |
Clearance Requirements
To ensure proper service access and ventilation, provide the CELL-DYN Ruby with the clearances shown in the following table.Clearance Requirements
Unit | Above | Behind | Left | Right |
---|---|---|---|---|
Analyzer | 15.2 cm (6 in.) | 15.2 cm (6 in.) | 15.2 cm (6 in.) | 15.2 cm (6 in.) |
Unit | Above | Behind | Left | Right |
---|---|---|---|---|
Analyzer | 30.5 cm (12 in.) | 15.2 cm (6 in.) | 40.6 cm (16 in.) | 40.6 cm (16 in.) |
Waste Disposal Requirements
All waste produced by the CELL-DYN Ruby must be disposed of according to local, state, and federal regulations governing the treatment and disposal of medical waste. Label all waste containers as biohazardous waste.Operating Noise Level and Heat Output
The CELL-DYN Ruby produces a certain amount of noise and heat as a normal part of operation. The following levels of noise and heat can be expected:Noise Level : | Idle Mode < 60 db Running Mode < 65 db |
|||
Heat Output : (non-condensing) Indoor Use |
0.6 kW maximum (2000 BTU) |